TRACKSENSE® PRO TECHNICAL SPECIFICATIONS

Multi Reader Station

Operating Temperature

5°C to 40°C/ 41°F to 104°F

Operating Voltage

5V External Adapter 100-240V AC, 50-60 Hz or USB

Reader/Logger Communication

Induction

PC Interface

USB

Capacity

1-4 Loggers at a Time

Expansion

1-3 Modules

Expansion Capacity

5-16 Loggers

Single Reader Station

Operating Temperature 5°C to 40°C/ 41°F to 104°F                               
Operating Voltage NA
Reader/Logger Communication Induction
PC Interface USB
Capacity 1 Logger at a Time
 

TSP Logger

Logger Basic Pro Pro X Pro XL
Operating Temperature: -30 to +105°C

-50 to +150°C

-80 to +150°C

-80 to +150°C

Operating Pressure: *

0 mbar to 10 bar abs

0 mbar to 10 bar abs

0 mbar to 10 bar abs

0 mbar to 10 bar abs

Material:

316L Stainless Steel

316L Stainless Steel

316L Stainless Steel

316L Stainless Steel

Diameter:

25 mm.

25 mm.

25 mm.

25 mm.

Length:

44 mm.

44 mm.

44 mm.

68.8 mm.

Weight With Battery:

48 Grams

48 Grams

48 Grams

-

Memory Capacity:

60,000 Data Points

60,000 Data Points

60,000 Data Points

60,000 Data Points

Minimum Sample Rate:

1 Second

1 Second

1 Second

1 Second

Maximum Sample Rate: 24 Hours

24 Hours

24 Hours

24 Hours

Maximum Start Delay: 14 Days

14 Days

14 Days

14 Days

Intrinsically Safe: Ex II G 
EEx ia IIC T3
-30 to +105°C

Ex II G 
EEx ia IIC T3
-50 to +105°C

Ex II G 
EEx ia IIC T3
-60 to +105°C

-

Compliance: CE

CE

CE

CE

Sensor: Interchangeable

Interchangeable

Interchangeable

Interchangeable

Time Accuracy: ±5 Seconds 
Per 24 Hours

±5 Seconds 
Per 24 Hours

±5 Seconds 
Per 24 Hours

±5 Seconds 
Per 24 Hours

Battery: TSP Standard Battery TSP Standard Battery TSP Standard Battery TSP 150L Battery
*In accordance with UL Certificate 06 ATEX 140022X, group 2, TrackSense® Pro, Pro X and Pro Basic loggers equipped 
  with pressure sensors and relative humidity sensors can not be exposed to more than 1.1 bar and 105°C when 
  working in an environment where intrinsically safe equipment is required. 
 

Technical Specifications for rigid sensors

 

Type

                  Rigid SS Standard

                       Rigid SS High

Diameter

                2 mm / 3 mm

                      3 mm

Length

                 From 10 mm

                From 150 mm 

Dimension of sensing element

                1.2 x 1.6 mm

                 1.2 x 1.6 mm

Position of measuring point:

 

 

From round tip

                   3 mm

                       3 mm

From sharp tip

                   6 mm

                       6 mm 

From conical tip

                   8 mm

                       8 mm

Temperature measuring range

            -196°C to +150°C

                   0 to +400°C

Accuracy:

 

 

Low -196 to -80°C

                  ± 0.5°C

                        -

Low -80 to -50°C

                  ± 0.3°C

-

Low -50 to -40°C

                  ± 0.2°C

-

Low -40 to -25°C

                  ± 0.1°C

-

Low -25 to 0°C

                  ± 0.05°C

-

Standard 0 to 100 °C

                  ± 0.05°C

-

Standard +100 to +140°C

                  ± 0.05°C

-

Standard +140 to +150°C        

                  ± 0.05°C

-

High 0 to +400°C

                        -

                    ± 0.5°C

Sensor capacity

1 or 2     

 1

Sensor response time:

 

 

T-10%

0.11 sec. / 0.22 sec.

 

T-50%

0.35 sec. / 0.72 sec.

 

T-63%

0.5 sec. / 1.00 sec.

 

T-90%

1.10 sec. / 2.00 sec.

 

Pressure operating range

0.001 mBar to 10 Bar ABS 

0.001 mBar to 10 Bar ABS 

 

Technical Specifications for Stainless Steel sensors

 

Type

Semi Flexible SS Standard

Semi Flexible SS High

Diameter

1.5 mm / 2 mm / 3 mm

 2.0 mm / 2.5 mm

Length

100 -1000 mm

150 - 1000 mm

Dimension of sensing element

 0.8 x 3.0 mm / 1.2 x 1.6 mm / 1.2 x 1.6 mm

 1.2 x 1.6 mm

Position of measuring point:

 

 
From round tip

3 mm

3 mm

From sharp tip

6 mm

6 mm 

From conical tip

8 mm

8 mm

Temperature measuring range

 -196°C to +150°C

 0 to +400°C

Accuracy:    
Low -196 to -80°C

 ± 0.5°C

-

Low -80 to -50°C

± 0.3°C

-
Low -50 to -40°C

± 0.2°C

-
Low -40 to -25°C

± 0.1°C

-
Low -25 to 0°C

± 0.1°C

-
Standard 0 to 100°C

± 0.1°C

-
Standard +100 to +140°C

± 0.1°C

-
Standard +140 to +150°C

± 0.1°C

-
High 0 to +400°C

-

± 0.5°C

Sensor capacity 1 or 2 1 or 2
Sensor response time:    
T-10% - / 0.11 sec. / 0.22 sec.  -
T-50% - / 0.35 sec. / 0.72 sec.  -
T-63% 0.46 sec. / 0.41 sec. / 1.00 sec.  1.46 sec. / 6.96 sec.
T-90% 0.67 sec. / 0.81 sec. / 2.00 sec.  2.87 sec. / 17.72 sec.
Pressure operating range 0.001 mBar to 4 Bar ABS  0.001 mBar to 4 Bar ABS 

 

Technical Specifications for Flexible sensors

 

Type

Flexible Low

Flexible Standard

Diameter

1.2 mm

1.8 mm

Length

100 - 500 mm

100 - 1000 mm

Dimension of sensing element

 0.8 x 3 mm

 1.2 x 1.6 mm

Position of measuring point:

 

 
From round tip

3 mm

3 mm

Temperature measuring range

 -196°C to +100°C*

 -196°C to +140°C

Accuracy:    
Low -196 to -80°C

± 0.5°C

± 0.5°C

Low -80 to -50°C

± 0.3°C

± 0.3°C

Low -50 to -40°C

± 0.2°C

± 0.2°C

Low -40 to -25°C

± 0.1°C

± 0.1°C

Low -25 to 0°C

± 0.1°C

± 0.1°C

Standard 0 to 100°C

± 0.1°C

± 0.1°C

Standard +100 to +140°C

-

± 0.1°C

Standard +140 to +150°C

-

-
High 0 to +400°C

-

-

Sensor capacity 1, 2 or 4 1 or 2
Sensor response time:    
T-10%  -  -
T-50%  -  -
T-63%  1.34 sec.  2.16 sec.
T-90%  2.55 sec.  4.31 sec.
Pressure operating range 0.001 mBar to 1 Bar ABS 

0.001 mBar to 4 Bar ABS 

 

*Can be sterilized at 125°C
 

TSP Pressure Sensors

  Pressure Sensor
Material 316L Stainless Steel
Measuring Principal Strain Gauge
Operating Temperature 0 to +150°C
Operating Pressure 10 mBar to 10 Bar Absolute
Measuring Range 10 mBar to 6 Bar Absolute, Up To 7 Bar Can Be Ordered         
Resolution 1 mbar
Measuring Accuracy ±0.25% Full Scale
 

TSP Combined Pressure / Temperature

  General Sensor Specifications
Material 316L Stainless Steel
Operating Temperature      0 to +150°C
Sampling 1 Sec. To 24 Hours Auto Start/Stop                                                

 

  Pressure Sensor
Measuring Principal        Strain Gauge
Operating Pressure 10 mBar to 10 Bar Absolute
Measuring Range 10 mBar to 6 Bar Absolute, Up To 7 Bar Can Be Ordered       
Resolution 1 mbar
Measuring Accuracy ±0.25% Full Scale

 

 

  Temperature Sensor                                                                
Measuring Principal External PT1000                                                                             
Length Of Sensing Element 20 mm
Measuring Range 0 to +150°C
Resolution 0.012°C
Measuring Accuracy ±0.05°C

 

 

Relative Humidity Sensors

 

1. Stability Chambers*

2. EtO/Stability

3. EtO

RH Sensing Element:

Capacitance

Capacitance

Capacitance

Temperature Sensing Element:

Chip RTD

PT1000

PT1000

 

 

   
Relative Humidity:      

Range (-40 to +80°C):

0-100%

0-95%

0-100%

Resolution:

0.10%

0.10%

0.10%

Response Time 
(63% In Slowly Moving Air):

8 Sec.

6 Sec.

50 Sec.

Accuracy At 25°C Non-Condensing:

     

(10% to 90%):

±2%

±2%

±2%

 

 

   
Temperature:      
Range:

0 to +90 °C

0 to +90 °C

0 to +90 °C

Resolution:

0.01°C

0.01°C

0.01°C

Response Time:

Max 30 Sec.

 Max 5 Sec.

Max 30 Sec.

Accuracy:      

+5°C to +40°C

±0.5°C

±0.1 oC

±0.1°C

+40°C to +65°C

±1.0°C

±0.1 oC

±0.1°C

+65°C to +90°C

±1.5°C

±0.1 oC

±0.1°C

* Requires ValSuite 2.4 or newer
 

Conductivity Sensor

Conductivity range:

 

0 to 200 μS/cm or 200 to 2000 μS/cm

Resolution:

0.1 μS/cm

Accuracy:

± 1 μS/cm or ± 2.5 % of reading whichever is worse

Temperature Range:

0°C to 100°C *

Accuracy: ± 0.05°C

Contact Material:

Peek (FDA compliant)

IP Class:

IP67

Temperature Sensor:

Pt-1000 to provide automatic temperature compensation

Dimensions L x D:

 37 mm x 25 mm / 13 mm

Immersion Depth:

Minimum 25 mm

*The Conductivity sensor is not designed for steam sterilization, but can be autoclaved at 121°C.